FDA Recall
Terminated
Historical recordClass II
L-Glutathione 200 mg/mL PF, 100 mL, Medaus Pharmacy 1-800-526-9183
by Medaus, Inc.
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Taurine 50 mg/mL PF, 100 mL, Medaus Pharmacy 1-800-526-9183
by Medaus, Inc.
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Lipo-Injection w/ Lidocaine (PF), Ascorbic Acid 50 mg; B1-50 mg; B2-5-PO4-5 mg; B3-50 mg; B5-5 mg; B6-5 mg; Cyano B12-100 mcg; Methionine 12.5 mg; Inositol-25 mg; Choline-25 mg; Lidocaine 10 mg/mL, 30 mL, Medaus Pharmacy 1-800-526-9183
by Medaus, Inc.
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Testosterone Cyp 200 mg/mL, W/Pres. (Benzyl ETOH), 3 mL, Medaus pharmacy 1-800-526-9183
by Medaus, Inc.
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Epinephrine 1:1,000 (1mg/mL) Sterile Solution for injection, Preservative Free/Sulfite Free, 1-mL Single-Use Syringe, Sterile Solution For Office Use Only, JCB Laboratories, Wichita, KS.
by JCB Labs LLC
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Sodium Citrate 4% Solution for Injection, 30 mL Multi Dose Vial, Sterile Solution for Office Use Only, JCB Laboratories, Wichita, KS.
by JCB Labs LLC
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Sodium Citrate 4% containing Gentamicin 320 mcg/mL, Solution for Injection, 30 mL Multi Dose Vial, Sterile Solution For Office Use Only, JCB Laboratories, Wichita, KS.
by JCB Labs LLC
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Sodium Thiosulfate 25%, 250 mg/mL (12.5 g/50mL), PF Sterile Solution for Injection, 50 mL Single Dose Vial, For Office Use Only, JCB Laboratories, Wichita, KS.
by JCB Labs LLC
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Acetylcysteine 20% Solution, Preservative Free, 4 mL Sterile Single Dose Vial, Not for Injection, Sterile Solution For Office Use Only, JCB Laboratories, Wichita, KS.
by JCB Labs LLC
- Reason
- Lack of Assurance of Sterility: The product lots are being recalled due to laboratory results (from a contract lab) indicating microbial contamination. The FDA was concerned test results obtained from the recalling firm's contract testing lab may not be reliable. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
IBU (TM) IBUPROFEN Tablets, USP, 800 mg, 500 -count bottle,(NDC 55111-684-05) Rx only, Distributed by Dr. Reddy's Laboratories, Inc., Bridgewater, NJ 08807 USA.
by Dr. Reddy's Laboratories, Inc.
- Reason
- Labeling: Incorrect or missing lot and/or exp date- This product is being recalled due to an incorrect expiration date of 05/2017. The correct expiration date is 10/2016.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Lidocaine/Phenylephrine PF 1%/1.5% Injectible vials, 1 mL Single Use Vial, Compounded by Leiter's Compounding Pharmacy, 1700 Park Ave., # 30, San Jose, CA 95126
by Leiter's Pharmacy
- Reason
- Lack of Assurance of Sterility: Leiter Pharmacy is recalling due bacterial contamination found after investigation at contract testing laboratory discovered that OOS/failed results were reported to customers as passing. Hence the sterility of these products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Bevacizumab (Avastin) 2.5mg/0.1mL PF Inj., 0.1 mL syringe, Compounded by Leiter's Compounding Pharmacy, 1700 Park Ave., # 30, San Jose, CA 95126
by Leiter's Pharmacy
- Reason
- Lack of Assurance of Sterility: Leiter Pharmacy is recalling due bacterial contamination found after investigation at contract testing laboratory discovered that OOS/failed results were reported to customers as passing. Hence the sterility of the products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Hyaluron (hyaluronidase) 150 units/mL Preserved Solution, 1 mL, Multi Dose Vial, Compounded by Leiter's Compounding Pharmacy, 1700 Park Ave., # 30, San Jose, CA 95126
by Leiter's Pharmacy
- Reason
- Lack of Assurance of Sterility: Leiter Pharmacy is recalling due bacterial contamination found after investigation at contract testing laboratory discovered that OOS results were reported to customers as passing. Hence the sterility of these products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
Dextrose 50% Preservative Free SDV Injection Solution, 50 mL Single dose vial, IV Administration, Compounded by Leiter's Compounding Pharmacy, 1700 Park Ave., # 30, San Jose, CA 95126
by Leiter's Pharmacy
- Reason
- Lack of Assurance of Sterility: Leiter Pharmacy is recalling due bacterial contamination found after investigation at contract testing laboratory discovered that OOS results were reported to customers as passing. Hence the sterility of these products cannot be assured.
- Published
- September 25, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
SODIUM HYDROXIDE 10% (W/V)/WATER FOR INJECTION, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
SODIUM BICARBONATE PF 8.4% INJECTABLE, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
SERMORELIN ACETATE 10 MG INJECTABLE, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
SELENIUM 400 MCG/ML INJECTABLE, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
QUAD AMINO BLEND 150/75/150/75 MG/ML INJECTABLE, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →
FDA Recall
Terminated
Historical recordClass II
PYRIDOXINE HCL P/F 100 MG/ML INJECTABLE, all presentations, FOR INTRAVENO AND SQ USE ONLY, STERILE, RX ONLY, Compounded by Beacon Hill Medical Pharmacy PC, dba Rxtra Solutions, Southfield, MI
by Beacon Hill Medical Pharmacy, P.C.
- Reason
- Lack of Assurance of Sterility; FDA inspectional findings resulted in concerns associated with quality control procedures that impacted sterility assurance
- Published
- September 18, 2013
- Distribution
- Nationwide
View recall →